Including Adapter; Gas Dryer Line; Gas Sampling Line; Nasal Sampling Line; T-fitting Tube. Product Description The ETCO2 module is the third ETCO2 version of National Medical. The new design is easier to match with monitor or anesthesia machine, will make maintenance much convenient. The CO2 sensor output parameters include real time CO2 concentration, respiration rate, end-tidal CO2, inspired CO2, The CO2 sensor module includes a miniature CO2 sensor, barometric pressure transducer, sampling flow control and a miniature low-power vacuum pump. The CO2 sensor can measure atmospheric pressure automatically, and has pressure compensated function. The host sends commands and receives waveform, breath parameters and command responses on two asynchronous serial data lines (TTL or RS232 level). Specification: Principle of Operation: Non-dispersive infrared (NDIR)/ InfraRed Spectroscopy; Initialization Time: Capnogram displayed in less than 20 seconds, Reaching full specifications within 2 minutes; CO2 Measure Range: 0–150 mmHg/0–19.7 %/0–20 kPa; CO2 Accuracy: 0-40mmHg ±2 mmHg 41- 70 mmHg ±5% of reading 71 – 100mmHg ±8% of reading 101-150mmHg ±10% of reading; CO2 Measure Resolution: 0.1mmHg; RR Measure Range: 3 - 150 BPM RR Accuracy: ±1BPM; Calibration: Don’t need user often calibration; Compensation parameters: Barometer Pressure 450 - 850 mmHg.
Trustpilot
1 month ago
3 days ago